Acquisition characteristics accuracy, Common-mode characteristics – National Instruments NI 5911 User Manual

Page 26

Advertising
background image

Appendix A

Specifications

© National Instruments Corporation

A-3

NI 5911 User Manual

Acquisition Characteristics

Accuracy

DC gain accuracy ................................... ±0.05% signal ±0.0001% fs

for all input ranges at 1 MS/s in
flexible resolution mode

DC offset accuracy................................. ±0.1 mV ±0.01% fs

for all input ranges at 1 MS/s in
flexible resolution mode

Input coupling ........................................ DC and AC, software selectable

AC coupling cut-off frequency
(–3 dB) ................................................... 2.3 Hz ±13%

Input impedance ..................................... 1 M

Ω ±2%

Max measurable input voltage ............... ±10 V (DC + peak AC)

Input protection ...................................... ±42 VDC (DC + peak AC)

Input bias current ................................... ±1 nA, typical at 25 °C

Common-Mode Characteristics

Impedance to chassis ground ................. 10 k

Common-mode rejection ratio ............... CMRR > –70 dB, (F

in

< 1 kHz)

±0.5 V

148 dBfs/

±0.2 V

140 dBfs/

±0.1 V

134 dBfs/

Input Range

Noise Referred to Input

Hz

Hz

Hz

Advertising