Appendix A

Specifications

NI 5911 User Manual

A-8

ni.com

Operating Environment

Note

Multiple NI 5911s in the same computer may raise operating temperatures beyond

specification and give rise to imprecise data. NI strongly recommends leaving an empty
PCI slot between multiple NI 5911s or adding a fan.

Ambient temperature ..............................5 to 40 °C

Relative humidity ...................................10% to 90%, noncondensing

Storage Environment

Ambient temperature ..............................–20 to 65 °C

EMC Compliance

CE2001, FCC

Calibration

Internal....................................................Internal calibration is done upon

software command. The
calibration involves gain, offset
and linearity correction for all
input ranges and input modes.

Interval.............................................1 week, or any time temperature

changes beyond ±5 °C. Hardware
detects temperature variations
beyond calibration limits, which
can also be queried by software.

External...................................................Internal reference requires

recalibration

Interval.............................................1 year

Warm-up time.........................................1 minute