Appendix a specifications, Specifications, Acquisition system – National Instruments NI 5911 User Manual

Page 24

Advertising
background image

© National Instruments Corporation

A-1

NI 5911 User Manual

A

Specifications

This appendix lists the specifications of the NI 5911. These specifications
are typical at 25

°C unless otherwise stated.

Acquisition System

Bandwidth .............................................. 100 MHz maximum,

see Table 2-2,

Available

Sampling Rates and
Corresponding Bandwidth in
Flexible Resolution Mode

Number of channels ............................... 1

Number of flexible resolution ADC....... 1

Max sample rate ..................................... 1 GS/s repetitive,

100 MS/s single shot

Resolution

Sample Rate

Mode

Effective Resolution

100/n

*

MS/s

Oscilloscope

8 Bits

12.5 MS/s

Flexible Resolution

11 Bits

5 MS/s

Flexible Resolution

14 Bits

2.5 MS/s

Flexible Resolution

15.5 Bits

1 MS/s

Flexible Resolution

17.5 Bits

500 kS/s

Flexible Resolution

18 Bits

200 kS/s

Flexible Resolution

18.5 Bits

100 kS/s

Flexible Resolution

19 Bits

50 kS/s

Flexible Resolution

19.5 Bits

Advertising