Figure 3-4. below-low-level analog triggering mode, E 3-3, Es 3-3 – National Instruments NI 6120 User Manual

Page 30

Advertising
background image

Chapter 3

Hardware Overview

NI 6115/6120 User Manual

3-6

ni.com

Figure 3-3. Analog Trigger Block Diagram for the NI 6115/6120

In below-low-level analog triggering mode, the trigger is generated when
the signal value is less than lowValue, as shown in Figure 3-4. HighValue
is unused.

Figure 3-4. Below-Low-Level Analog Triggering Mode

In above-high-level analog triggering mode, the trigger is generated when
the signal value is greater than highValue, as shown in Figure 3-5.
LowValue is unused.

PGIA

Analog

Input

CH0

+

ADC

ADC

ADC

DAQ-STC

Analog
Trigger

Circuit

Mux

PGIA

Analog

Input

CH1

+

PGIA

Analog

Input

CH2

+

PGIA

Analog

Input

CH3

+

ADC

PFI0/TRIG1

Digital Data

AC Couple

10 k

Trigger

DAC

Trigger

DAC

highValue

lowValue

lowValue

Trigger

Advertising