Performed or Supported Signal Conditioning

This section provides information on types of signal conditioning
performed by the SCXI-1321 or is supported by it.

Offset Nulling

Offset nulling is a hardware nulling procedure used with
Wheatstone-bridge transducers that have an initial offset error. Correcting
this error improves measurement accuracy. The nulling circuitry operates
with quarter-bridge, half-bridge, and full-bridge strain-gauge
configurations. Each channel has its own nulling circuitry and its own
trimming potentiometer as listed in Table 1.

To null the static offset voltage of the bridge, complete the following steps:

1.

Connect the bridge configuration to the selected channel.

2.

Select and read the channel output.

3.

While monitoring the output, rotate the trimmer wiper with a flathead
screwdriver until you reach 0 V.

You have nulled the bridge and are ready for a measurement.

The nulling range that is provided with the terminal block is ±2.5 mV,
assuming that you have a 120

Ω strain gauge quarter-bridge configuration

and 3.3 V excitation voltage. You can change this range by replacing the
nulling resistor with a resistor of another value. Each channel has an
independent socketed nulling resistor. Therefore, you can mix the ranges to
accommodate each channel requirement. Table 2 lists the nulling resistors
and their corresponding channels.

Table 1. Trimmer Potentiometers and Corresponding Channels

Channel Number

Trimmer Potentiometer

0

R1

1

R2

2

R14

3

R15