Chapter 2 hardware overview, Figure 2-1. ni 5911 block diagram, Differential input – National Instruments NI PCI-5911 User Manual

Page 15: Differential input -1, Hardware overview

Advertising
background image

© National Instruments Corporation

2-1

NI PCI-5911 User Manual

2

Hardware Overview

This chapter includes an overview of the NI 5911, explains the operation of
each functional unit making up the NI 5911, and describes the signal
connections. Figure 2-1 shows a block diagram of the NI 5911.

Figure 2-1. NI 5911 Block Diagram

Differential Programmable Gain Input Amplifier (PGIA)

The analog input of the NI 5911 is equipped with a differential
programmable gain input amplifier. The PGIA accurately interfaces to and
scales the signal presented to the ADC regardless of source impedance,
source amplitude, DC biasing, or common-mode noise voltages.

Differential Input

When measuring high dynamic range signals, ground noise is often a
problem. The PGIA of the NI 5911 allows you to make noise-free signal
measurements. The PGIA differential amplifier efficiently rejects any
noise present on the ground signal. Internal to the PGIA, the signal

Protect/

Calibration

Mux

Calibration

Generator

PGIA

Noise

Shaper

Reference
Clock

Data

A/D Converter

100 MHz, 8-Bit

Digital Signal

Processor

Digital I/O

Connector

Analog Input

Connector

AC/DC Coupling

Capture

Memory

Timing I/O,

Memory Control

Advertising