NI 9225E Operating Instructions and Specifications

Isolation Voltages

Channel-to-channel

Continuous ................................. 600 V

rms

,

Measurement Category II,
(Basic insulation)

Withstand.................................... 2,300 V

rms

, verified by a 5 s

dielectric withstand test

Channel-to-earth ground

Continuous ................................. 300 V

rms

,

Measurement Category II,
(Double insulation)

Withstand.................................... 2,300 V

rms

, verified by a 5 s

dielectric withstand test

Measurement Category II is for measurements performed on
circuits directly connected to the electrical distribution system.
This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S.
or 230 V for Europe.

Caution

Do not connect the NI 9225E to signals or use for

measurements within Measurement Categories III or IV.