Appendix C

Error Codes and Solutions

© National Instruments Corporation

C-3

NI-488.2 User Manual

Linux

Enter the following command:

<InstallDir>/natinst/ni4882/bin/gpibtsw

where

<InstallDir>

is the directory where you chose to install NI-488.2.

The default is

/usr/local

.

Use the unit descriptor returned from

ibdev

or

ibfind

as the first

parameter in subsequent traditional NI-488.2 calls. Examine the
variable before the failing function to make sure its value has not been
corrupted.

ECIC (1)

ECIC is returned when one of the following functions is called while the
interface is not CIC:

Any device-level traditional NI-488.2 calls that affect the GPIB.

Any board-level traditional NI-488.2 calls that issue GPIB command
bytes:

ibcmd

,

ibcmda

,

ibln

, and

ibrpp

.

ibcac

and

ibgts

.

Any NI-488.2 multi-device calls that issue GPIB command bytes:

SendCmds

,

PPoll

,

Send

, and

Receive

.

Solutions

Possible solutions for this error are as follows:

Use

ibsic

or

SendIFC

to make the GPIB interface become CIC

on the GPIB.

Use

ibrsc

1

to make sure your GPIB interface is configured as

System Controller.

In multiple CIC situations, always be certain that the CIC bit appears
in the status word

ibsta

before attempting these calls. If it does not

appear, you can perform an

ibwait

(for CIC) call to delay further

processing until control is passed to the interface.