Chapter 4 SelfCalibration

Setting the calibration schedule [Schedule]

Set the calibration schedule for SelfCalibration.

Attention

• Clock adjustment is necessary for SelfCalibration to operate.

Procedure
1. Choose <Setting> from the SelfCalibration menu, and press

.

2. Choose <Schedule> from the Setting menu, and press

.

The schedule setting menu is displayed.

3. Select the items to change with

or

, and press

.

Setting Item

Setting Range

Setting

Starting Time Off

Power Save

Immediately

Set the timing at which SelfCalibration is executed.

Off

: SelfCalibration is not executed.

Power Save : When the time period set in the schedule

has elapsed, it is executed if the mode

initially switches to power save, or if the

power initially switches off.

Immediately : When the time period set in the schedule

has elapsed, it is executed immediately.

Setting Type

Type1

Type2

Type3

Set according to the frequency at which SelfCalibration is

executed.

Type1 : Once every few months

Type2 : Once every few weeks

Type3 : Once every few hours (monitor usage time)

Interval

Type1 : 1 - 12 months

Type2 : 1 - 5 weeks

Type3 : 50 - 500 h

Set the interval between the completed SelfCalibration and

the next execution. The setting details differ depending on

the type of <Setting Type> specified.

Week

1st week - 5th week

Set which week to execute SelfCalibration on for the month

in which calibration is executed.

Day

MON/TUE/WED/THU/

FRI/SAT/SUN

Set which day to execute SelfCalibration on for the week in

which calibration is executed.

Time

00:00 - 23:30

Set the time at which SelfCalibration is executed.

*1

Settings that are used frequently can be preset.

4. Adjust using

or

.

5. Press

to exit the adjustment.

The calibration schedule is set.

NOTE

• The schedule can be set using ColorNavigator. Refer to the ColorNavigator User’s Manual on the CD-ROM.
• When the starting time is [Immediately], and the setting type is either [Type1] or [Type2], SelfCalibration is

executed according to schedule even if the power is switched off using

.

• When the starting time is “Off”, the SelfCalibration menu shows “Setting Incomplete.” However, SelfCalibration

can be executed manually.