IX81-ZDC2

11

Put the specimen in the glass-bottom dish (with cover glass thickness

of 0.16 to 0.18 mm) and place the dish on the center of the stage.

} If the dish is prone to slide on the stage, attach the stage clips (IX-SCL)

1 or use the provided weight.

1

1. Run the “Application Software for PC.”
2. Set up the AF (autofocusing) functions.
The following parameters should be set to start AF.
In one-shot mode

· Offset : Setup of the amount of compensation between the cover glass

top surface and target position.

· AF zone : AF search zone (AF center, width and range)

In continuous focus mode
Confirm that the cover glass top surface is detected in the one-shot

mode before switching to the continuous focus mode.

} For details, refer to the instruction manual and Help of the application.

When using the MetaMorph application software, install IX2-

BSW Ver. 1.7 (in the CD provided with this product) in your PC
in advance.