Ix81-zdc2 – Olympus IX81-ZDC2 User Manual

Page 7

Advertising
background image

4

IX81-ZDC2

5

3. Applicable objectives:

· If objective has an aperture, set it wide open.
·
The AF focusing speed is approximately 1 to 5sec. in the focusing range,

but slows down depending on the AF search range setting.

· The focus drift compensation is not available with the phase contrast

objectives because these incorporate a phase plate.

· The continuous focus mode can be used only with an oil- or water-immersion

objective. If an objective other than oil- or water-immersion objective is used,
the one-shot focus mode should be used.

· When using a large specimen such as a micro-test plate, focusing may not

be achieved due to its inclination.

The focusing is guaranteed only with a glass bottom dish with a diameter

of 35 mm.

Objective Name

NA

WD (mm)

APON 60X O

TIRF

UAPON 100X O

TIRF

1.49

1.49

0.1

0.1

PLAPON 60X O

1.42

0.15

UPLSAPO 60X O

UPLSAPO 100X O

1.35

1.40

0.15

0.13

UPLFLN 100X O2

UPLFLN 100X O

I2

1.30

1.30-0.60

0.20

0.20

UPLSAPO 20X

*

UPLSAPO 40X2

*

UPLSAPO 60X W

0.75

0.95

1.20

0.60

0.18

0.28

UPLFLN 20X

*

UPLFLN 40X

*

UPLFLN 60X

*

0.50

0.75

0.90

2.10

0.51

0.20

LUCPLFLN 20X

*

LUCPLFLN 40X

*

LUCPLFLN 60X

*

0.45

0.60

0.70

6.60-7.80

2.70-4.00

1.50-2.20

UAPO 40X O

I3/340

1.35-0.65

0.1

* Compatible with one-shot focus mode only.

} The previous models to the objectives listed above (the objectives having a smaller number or no number at the end of

model name) can also be used.

4. With DIC (Differential Interference Contrast) observation, focus compensation is possible only with the gray sensitive color

and in the one-shot focus mode. It is neither possible in phase contrast observation.

5. Operating environmental temperatures: 18 to 40°C (64.4 to 104°F)

(Focus compensation is possible under a temperature variation range within
5°C (41°F) and when the microscope is protected against the wind from air
conditioners.)

CAUTION