Piping Precautions

(1) Flange connection

The flowmeter has upstream and downstream

flanges on the ends of detector pipe. Connect

these flanges with the flanges on both sides of

pipeline bore using connection bolts after

inserting a gasket between them. See Figure 7.

Tighten the bolts in even increments diagonally

across.

Figure 7. Flange Connection

(2) Required straight pipe length

The straight pipe length is required to prevent

uneven flow velocity and a disturbance on the

fluid surface in the detector pipe. The required

straight pipe length should be as follows:

• Downstream side:L = 5D minimum

• Upstream side:L = 10D minimum

where,

L = straight pipe length (straight pipeline length

+ half length of detector pipe length)

Note: The length of reducers, if used, can be counted as

a part of straight pipe length.

Figure 8. Required Straight Pipe Length on

Upstream and Downstream Sides

About establishment environment

Do not store or install the flowmeter:

• Where there is direct sunlight.
• Where excessive vibration or mechanical shock

occurs.

• Where high temperature or high humidity

conditions exist.

• Where corrosive atmospheres exist.
• Places that can be submerged under water.
• Where there is sloped floor. To put the flowmeter

temporarily on the floor, place it carefully with

something, such as stopper, to support it so that the

flowmeter will not topple over.

In areas like the following, there may be the case that
infrared switches do not function correctly. (If these
are unavoidable, use an appropriate cover.)
(1) Where unit (operation panel) is exposed to direct

sunlight, reflection of light onto window pane and
diffused light reflection.

(2) Where smoke and steam may occur.

(3) Where exposed to direct snow, ice or mud.

Ordering Information

1. When ordering the LF502 flowmeter, refer to

Table 1 (Type Specification Codes). An entry

must be made for each of the columns in each of

these tables.

2. Fluid characteristics:

(1) Type of fluid to be measured and its

characteristics

(2) Fluid temperature

(3) Fluid pressure

(4) Electrical conductivity of the fluid

3. Measuring range

4. Digital I/O specifications

5. Ordering scope:

Actual flow calibration data:(Required or not)

6. Other items

Specifications other than standard items

Observation

window

Observation

window

Upstream flange

Downstream flange

Flow

directio

Gasket

Bolt

Flow regulation

valve etc.

Observation window

Detector

10D minimum 5D minimum