24

Understanding.Out‑of‑Range.Control.Test.Results

WARNING

The.control.ranges.apply.only.to.control.results..They.only.indicate.that.your.test.strips.and.meter.

are.working.properly..Do.not.use.control.results.to.interpret.blood.glucose.results.

If the control result is not inside the acceptable range (printed on the test strip container), do not use the

meter until you solve the problem. Here are some things you can check:

Troubleshooting.Checks

Action

1. Are the test strips or control solution past the

“Use By” date or discard date?

If either is past the “Use By” date or discard date,

throw it away.

2. Did you wipe the tip of the control solution

bottle with a tissue prior to use?

Wipe the tip of the bottle with a tissue. Repeat the

control test with a new test strip and a fresh drop of

control solution.

3. Were the test strip container and control

solution bottle always closed tightly?

If you think either may have been uncapped for some

time, replace the test strips or control solution.

4. Was the test strip used immediately after

removing it from the container?

Repeat the control test with a new test strip.

5. Were the test strips and control solution stored

in a cool, dry place?

Repeat the control test with properly stored test strips

or control solution.