Accu-Chek Aviva Plus silver User Manual

Page 87

Advertising
background image

85

The code key is incorrect. Turn off

the meter and insert a new code key.

If this does not fix the problem,

contact the ACCU‑CHEK Customer

Care Service Center at

1‑800‑858‑8072.

Your blood glucose may be extremely

high or a meter or a test strip error

has occurred. If this confirms how

you feel, contact your healthcare

professional immediately. If it does

not confirm how you feel, repeat the

test and refer to Chapter 3, “Unusual

Test Results.” If this does not confirm how you feel,

perform a control test with your control solution and

a new test strip. If the control result is within the

acceptable range, review the proper testing

procedure and repeat your blood glucose test with a

new test strip. If the E‑3 code still appears for your

blood glucose test, your blood glucose result may be

extremely high and above the system’s reading

range.

Contact.your.healthcare.professional.

immediately. If the control result is not within the

acceptable range, see Chapter 2, “Understanding

Out‑of‑Range Control Test Results.”