CALIBRATION PROCEDURE
MODEL 1850
 
 
TOOLS REQUIRED: 
 
Straight Edge, Protractor, 3/32 Allen Wrench, Phillips Screwdriver 
 
 
 
INSTRUCTIONS: 
 
Turn power off 
Remove knee potentiometer cover 
Remove top cover from the unit 
 
Move SW1 slide switch located at the bottom left of the board from run to calibrate 
 
Using the up and down arrows move the unit to the zero position 
 
Use a straight edge to determine zero 
 
Look for these values on the hand control 
Ext. window (Letter F with a red dot)
ROM Window (2) Flexion Window (490) reading this as 2490 (2490 represents the zero setting)
 
If these values are not present then the unit is out of calibration 
 
If the above values do not appear on the hand control then it is necessary to move the knee pot to gain these values. To move 
the knee pot, loosen the clamps on the knee pot and turn the knee pot clockwise or counterclockwise to achieve the above 
readings. Once these values have been achieved then tighten the knee pot clamps. 
 
To lock in the zero (0) position, depress the ext. button 1 time and listen for an audible beep. 
 
By depressing the UP arrow button you can move the unit into the 90 degree (90) position, using a protractor or similar device, 
stopping at this position 
Once in the 90 degree position (90) and if you are in the approximate range a red dot will be visible in the flexion window. 
If the red dot is visible then depress the flexion button 1 time and listen for an audible beep. 
 
Switch unit from calibration mode to run mode 
Set hand control at -5 extension, 110 Flexion, 0 pause, and 10 speed 
 
Depress start/stop switch on the hand control and allow unit to travel through full range of motion. If the unit travels the full 
range of motion without binding then the unit has been calibrated successfully.