Remote Interface Reference

RS-232 Configuration

4.3 RS-232 Configuration

The Model 430 Programmer uses the following parameters related to the

RS-232 interface:

Baud Rate: 115200

Parity: No Parity

Data Bits: 8 Data Bits

Number of Start Bits: 1 bit

Number of Stop Bits: 1 bit

Flow Control: Hardware (RTS/CTS)

4.3.1

Serial Connector

An IBM-compatible computer’s serial port can be directly connected to the

Model 430 Programmer via a standard DB9-female-to-DB9-female null

modem serial cable. Refer to your computer’s documentation to determine

which serial ports are available on your computer and the required

connector type.

The Model 430 Programmer is classified as a DTE (Data Terminal

Equipment) device. It uses the standard DB9 male connector and identical

pinout used on IBM-compatible computers. The RS-232 connector pinout

for the Model 430 Programmer is fully documented on page 153 in the

Appendix.

4.3.2

Termination Characters

All commands and queries are transmitted and received as ASCII values

and are case insensitive. The Model 430 Programmer always transmits

<CR><LF> (a carriage return followed by a linefeed) at the end of an RS-

232 transmission. The Model 430 Programmer can accept <CR>, <LF>,

<CR><LF>, or <LF><CR>, or a semicolon ( ; ) as termination characters

from an external computer.

4.4 Ethernet Configuration

The Model 430 Programmer provides a 10/100Base-T Ethernet interface

as a standard feature. It complies with the IEEE 802.3u 100Base-TX and

802.3 10Base-T standards.

The Model 430 Programmer allows its IP address, subnet mask and

gateway IP address to be assigned either statically or dynamically. To

make these values static and assign them manually, set IP Address

Assignment to Static (see section 3.10.5.1) and then set the values using

Advertising