3 Operation

This section describes the operation of the Model 430 Programmer. Every

menu and submenu item is illustrated and described in detail. An example

setup of the Model 430 Programmer is presented in section 3.11 on

page 72. An example ramping operation is presented in section 3.14 on

page 85.

3.1

System Power On/Off Sequence

The Model 430 Programmer should always be energized before the power

supply that it is controlling. The Model 430 Programmer is designed to

prompt the user in order to ensure the power supply is energized at the

proper time. The Model 430 Programmer should always be de-energized

after the power supply is shut down.

3.1.1

Model 430 Programmer Power On/Off

Place the Model 430 Programmer power switch in the ON position. After

the Model 430 Programmer is powered on and fully initialized (about 20

seconds), the following display will appear:

After this screen is displayed, the power supply can be powered up (See

“Energizing Power Supply” on page 22.) followed by pressing the

ENTER

key on the Model 430 Programmer. This brings up the default display

1

.

Note

If turned off, the Model 430 Programmer must remain unpowered

for at least 5 seconds before it is powered back on. If not, there may

be an initialization error, in which case the following screen will be

displayed.

1. Refer to section 3.2 on page 22.

+0.00 A -

Turn on power supply

+0.00 Vs

Press ENTER to continue

AMI Model 430 Programmer
FAILURE TO LOAD.