3.3.3 User Calibration
User calibration is used when it is desired to calibrate the balance using a mass of known value. To use this calibra-
tion feature, proceed as follows:
(FLASHING)
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYPE is displayed.
• Press Enter button, CAL TYPE InCAL is dis -
played.
• Press
button twice to select USER calibration,
CAL TYPE USer is displayed.
• Press Enter button, the display indicates the last
calibration mass value which was entered with the
first digit flashing. (Sample illustrates 6000g).
• Press
and or
and enter the
desired mass value. This number must be at least
25% of the full span value.
• Press Enter button, WORKING is displayed.
NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.
• Display changes to PUT WT 6000g.
• Place specified calibration mass on pan.
• Press Enter button, WORKING is displayed. After a
few seconds CAL SET is displayed, the display then
returns to WEIGH mode.
User calibration is completed.
• Remove calibration mass from the pan.