5 percent weighing, Procedure – Ohaus E1M110 Explorer Balances User Manual

Page 22

Advertising
background image

22

3.5 Percent Weighing

Percent Weighing is

enabled only when Percent is turned ON. Percent weighing permits you to place a reference

load on the balance, then view other loads as a percentage of the reference. The load you place on the pan as a
reference may be displayed as any percentage you select from 5% to 100% (in 1% increments). One hundred percent
does not necessarily have to represent the reference load. Subsequent loads, displayed as a percentage of the
reference are limited only by the capacity of the balance. The default setting is Reference 100%. Refer to paragraph
4.8 to enable percent weighing.

NOTE: The PERCENT display (number of digits) is a
function of the accuracy of the balance and the size of
the reference weight. The display examples were with a
300g mass used with an 12 kilogram balance.

Procedure

• Remove the reference weight from the pan and re-

place it with another load. The second load is dis-

played as a percentage of the reference.

Balance displays reference weight for five seconds in
selected measuring unit, then displays the percentage.

• Press the Mode button.

• Press

or

button until PERCENT is dis-

played.

• Press Enter button, PUT>PAN 100% is displayed. If

a container is used, the balance can be tared at this

point.

• Put the reference load on the pan.

• Press

or

button and select reference

weight percentage (Percent Range 5 to 100). Hold

button down for fast change.

• Press Enter button to save setting, WORKING is

displayed ... calculating reference weight.

Advertising