7 animal weighing, Procedure – Ohaus E1M110 Explorer Balances User Manual
Page 24

24
Procedure
• Press the Mode button.
• Press
or
button until ANIMAL is displayed.
• Press Enter button to continue, LEVEL is displayed.
• Press
or
button to change animal weighing
level, 0, 1, 2 or 3. 0 level represents an inactive
subject, 3 is used for a very active subject.
• Press Enter button to continue, AUTO is displayed.
• Press
or
button to select AUTO ON or OFF.
• Press Enter button to continue.
3.7 Animal Weighing
Animal Weighing is
enabled only when Animal is turned ON in the Mode submenu.
When the AUTO function is set ON, different subjects can
be weighed one after another without pressing any
buttons. When the balance displays READY, simply
place subject on pan.
Start Animal Cycle
• Place animal container if used on pan.
• Press
to tare the container.
• Place subject on pan.
• The animal cycle will automatically start if AUTO was
set to ON.
• Press Enter button to start animal cycle if AUTO was
set to OFF.
Completed Animal Cycle
• Balance displays weight until specimen is removed
from the pan.
During Animal Cycle
• Display shows countdown to AW0.