Calbration menu, Calibration menu protection, Calibration masses – Ohaus GT4100DG Electronic Balances User Manual

Page 18: Calibration menu

Advertising
background image

18

NOTE

NOTE

NOTE

NOTE

NOTE: Cal Span, Cal Linearity
and Cal User are disabled for
Type Approved/LFT balances.

CALIBRATION MENU

CALIBRATION MENU

CALIBRATION MENU

CALIBRATION MENU

CALIBRATION MENU

Precision Advanced balances features CalTest

TM

which offers a choice of three

calibration methods: Cal Span, Cal Linearity, and Cal User.

Cal Span

Cal Span

Cal Span

Cal Span

Cal Span

calibration

ensures that the balance reads correctly within specifications using two weight values:
zero and a weight value at either 25%, 50%, 75% of or at the balance’s full capacity.

Cal Linearity

Cal Linearity

Cal Linearity

Cal Linearity

Cal Linearity

calibration minimizes deviation between actual and displayed weights

within the balance’s weighing range. Three weight values are used: zero, a weight
value at midpoint of the balances weighing range, and a weight value at or near the
balance’s specified capacity.

Cal User

Cal User

Cal User

Cal User

Cal User

is a method where the balance can be

calibrated using a mass of known value by entering that value into the balance.

Cal

Cal

Cal

Cal

Cal

Test

Test

Test

Test

Test

allows the stored calibration data to be tested against the current mass being

used for the test. The following figure illustrates the sequence in which submenus
appear on the Calibration menu. Item shown bolded is a default setting.

CALIBRATION MENU

CALIBRATION MENU

CALIBRATION MENU

CALIBRATION MENU

CALIBRATION MENU

CAL SPAN

0, 25%, 50%, 75% or Span Weight (100%

100%

100%

100%

100%).

CAL LINEARITY

0, Linearity & Span Weight.

CAL USER

Enter #

CAL TEST

0, Span Weight.

END, Calibration

Calibration Menu Protection

Calibration Menu Protection

Calibration Menu Protection

Calibration Menu Protection

Calibration Menu Protection

NOTES

NOTES

NOTES

NOTES

NOTES:

1. Calibration may be locked out to prevent unauthorized personnel from changing
calibration. If calibration has been locked out, you can only access Cal Test.

2. To lock out calibration menu, after calibration, refer to the section titled Menu
Lock-Out Protection.

Calibration Masses

Calibration Masses

Calibration Masses

Calibration Masses

Calibration Masses

A 2kg and 4 kg mass are required for calibration. If you begin calibration and realize
calibration masses are not available, exit the menu. The balance will retain previously
stored calibration data. Calibration should be performed as necessary to ensure
accurate weighing. Masses must meet or exceed ASTM Class 1 Tolerance. Calibra-
tion masses are available as accessories.