RS485 / RS422 Option Kit EN-3

• The RS485 / RS422 option may be configured for either duplex (4-wire) or half

duplex (2-wire) operating mode. Jumper W1 must be removed for duplex mode
and installed for half duplex mode.

Figure 2.3 Duplex Figure 2.4 Half Duplex

• A ground connection is provided on connector J2 for systems that require it.

A 120 Ohm termination resistance is provided by installing a jumper on connector

W2 of the RS485 / RS422 pc board. When W2 is shorted, the termination
resistance is placed between the RXD + and RXD - lines.


For more information about RS485 installations, refer to the document TSB-89-A

Application Guidelines for TIA/EIA-485-A

, published by the

Telecommunications

Industry Association.

.