Veris Industries 275 Manual User Manual

Page 10

Advertising
background image

19

18

Measuring Continuity

WARNING!

Do not attempt to make continuity measurements with circuit
energized. Make sure the temperature probe is NOT plugged
in during this test.

Instrument set-up:

FUNC.

BLACK RED
TEST LEAD

TEST LEAD

COM

V/

Measurement Procedure:

1.

Disconnect power to circuit to be measured.

2.

Plug black test lead into the COM input jack.

3.

Plug red test lead into V/

input jack.

4.

Set the rotary switch to the

position.

5.

Connect test leads to circuit to be measured.

6.

The 275 will beep and the LED will illuminate at
resistances of 35

or lower.

Optional Modes

(See Other Features Section)

D-H:

Freezes the reading on the LCD.

Measuring DC Amperage

CAUTION!

Do not attempt to make a current measurement with the
test leads. The 275 measures the current by clamping the
jaw around one conductor (wire). Clamping around more
than one wire will result in erroneous readings. Make sure
the temperature probe is NOT plugged in during this test.

Instrument set-up:

FUNC.

BLACK RED

MIN

MAX

TEST LEAD TEST LEAD

READING

READING

NOT USED NOT USED

0.01A

400A

Measurement Procedure:

1.

Disconnect power to circuit to be measured.

2.

Set rotary switch to

function.

3.

Clamp the jaws around one conductor of the circuit to
be measured. For best results, center the wire in the
jaw.

4.

Reconnect power to circuit to be measured.

5.

Read the current on the 275.

Optional Modes

(See Other Features Section)

D-H:

Freezes the reading on the LCD.

MAX/MIN:

Records minimum and maximum measure-

ments

P-H:

Displays peak amperage.

Advertising