With the menu item "
Operating range“ you
can define the operating range of the sensor
deviating from the "
Min/Max adjustment“. By
default, the operating range corresponds
otherwise to the min./max. adjustment (span).
Generally, it is better to choose the operating
range approx. 5 % bigger than the adjusted
measuring range (span) determined by the
min./max. adjustment.
In the example:
- Min. adjustment to 1.270 m,
- Max. adjustment to 5.85 m.
In the example you would have adjusted the
operating range from 1 m to 6 m.
• Save the adjustments and quit the menu
window "
Limitation of the operating range“.
• Click to
Measuring conditions.
• In the menu window "
Measuring conditions“
you click on the options corresponding to
your application.
• Confirm with
"OK.
After a few seconds of saving during which
the adjustments are permanently saved in
the sensor, you are again in the window
"
Meas. environment“.
Sonic velocity
When measuring in a gas composition devi-
ating from air, e.g. pure nitrogen, the sonic
velocity will change. Without correction, this
would distort the measuring result.
In the menu "
Sonic velocity“ the sonic velocity
can be adapted to the correct value in case
of another gas composition.
Setup - Adjustment with the PC on VEGAMET
• Confirm with "
OK“.
• Click in the menu window "
Meas. environ-
ment“ to
"Quit.
You are again in the menu window "
Sensor
optimisation“.