VEGASON 54V … 56V

With the menu item "

Operating range“ you

can define the operating range of the sensor
deviating from the "

Min/Max adjustment“. By

default, the operating range corresponds
otherwise to the min./max. adjustment (span).

Generally, it is better to choose the operating
range approx. 5 % bigger than the adjusted
measuring range (span) determined by the
min./max. adjustment.
In the example:
- Min. adjustment to 1.270 m,
- Max. adjustment to 5.85 m.
In the example you would have adjusted the
operating range from 1 m to 6 m.

• Save the adjustments and quit the menu

window "

Limitation of the operating range“.

• Click to

„Measuring conditions“.

• In the menu window "

Measuring conditions“

you click on the options corresponding to
your application.

• Confirm with

"OK“.

After a few seconds of saving during which
the adjustments are permanently saved in
the sensor, you are again in the window
"

Meas. environment“.

Sonic velocity

When measuring in a gas composition devi-
ating from air, e.g. pure nitrogen, the sonic
velocity will change. Without correction, this
would distort the measuring result.

In the menu "

Sonic velocity“ the sonic velocity

can be adapted to the correct value in case
of another gas composition.

Setup - Adjustment with the PC on VEGAMET

• Confirm with "

OK“.

• Click in the menu window "

Meas. environ-

ment“ to

"Quit“.

You are again in the menu window "

Sensor

optimisation“.