6 menu schematic, Basic adjustment flow, Basic setting level – VEGA VEGADIF 65 Profibus PA User Manual
Page 53
53
6 Adjustment with the display and adjustment module PLICSCOM
VEGADIF 65 • Profibus PA
36129-EN-130417
6.6 Menu schematic
Basic adjustment flow
▶
Basic adjustment
1
Display
Diagnostics
Service
Info
Sensor address
1.1
126
Applications
1.1
Flow
▼
Unit
1.1
Unit of measurement
bar
▼
Temperature unit
°C
▼
Position correction
1.2
Offset
DP
=
-0.0035 bar
0.0000 bar
Min. adjustment
1.3
000.0 %
DP
- +
=
0.0000 bar
0.0000 bar
Max. adjustment
1.4
100.00 %
DP
- +
=
0.5000 bar
0.0000 bar
Damping
1.5
1 s
Linearization curve
1.6
linear
▼
Leak flow volume
suppression
1.6
Activated
▼
Sensor-TAG
1.7
Sensor
Basic setting level
▶
Basic adjustment
1
Display
Diagnostics
Service
Info
Sensor address
1.1
126
Applications
1.1
Level
▼
Unit
1.1
Unit of measurement
bar
▼
Temperature unit
°C
▼
Position correction
1.2
Offset
DP
=
-0.0035 bar
0.0000 bar
Min. adjustment
1.3
000.0 %
=
0.0000 bar
0.0000 bar
Max. adjustment
1.4
100.00 %
=
0.5000 bar
0.0000 bar
Damping
1.5
1 s
Linearization curve
1.6
▶
Linear
Horizontal cylindrical tank
Spherical tank
User programmable
Sensor-TAG
1.7
Sensor