9 Maintenance and fault rectification

VEGADIF 65 • Profibus PA

36129-EN-130417

In Ex applications only one instrument and one electronics module

with respective Ex approval may be used.
The sensor serial number is required to order an exchange electron-

ics. You will find this number on the type label of the instrument or on

the delivery note.

9.4 Exchanging the electronics module

In case of a defect, the electronics module can be exchanged against

an identical type by the user. If no electronics module is available on

site, the module can be ordered from the agency serving you.
Ordering and exchange are possible with or without sensor serial

number. The electronics module with serial number includes order-

specific data such as factory setting, seal material etc. These are not

included in the electronics module without serial number.
The serial number is stated on the type label of VEGADIF 65 or on the

delivery note.

9.5 Software update

The software version of VEGADIF 65 can be determined as follows:

on the type label of the electronics

Via the display and adjustment module

via PACTware

You can view all software histories on our website www.vega.com.

Make use of this advantage and get registered for update information

via e-mail.
The following components are required to update the sensor soft-

ware:

Sensor

Voltage supply

VEGACONNECT

PC with PACTware

Current sensor software as file

At "www.vega.com/downloads" go to "Software". Select under "plics

sensors and instruments", "Firmware updates" the respective instru-

ment series and software version. Load the zip file via the right mouse

key with "Save target as" e.g. on the desktop of your PC. Move with

the right mouse key to the folder and select "Extract all". Save the

extracted files, for example on the desktop.

Connect the signal conditioning instrument to power supply and

provide the connection from the PC to the instrument via the interface

adapter. Start PACTware and go via the menu "Project" to the VEGA

project assistant. Select "USB" and "Set instruments online". Activate

the project assistant with "Start". The assistant establishes the con-

nection automatically and opens the parameter adjustment window

"Sensor # online parameter adjustment". Connect this parameter

adjustment window before you carry out further steps.

Load sensor software

to PC

Prepare update