For continuity tests, connect the test leads to two points of the circuit to

be tested. If continuity exists (resistance < 50Ω), the built-in buzzer will

sound. If the circuit is open, “OL” will be indicated.

Non-contact AC voltage test
•

Set the rotary switch [3] to any position (except OFF).

•

The AC-sensor is located at the top of the meter, near the light sensor

[9]. Point the AC-sensor towards the suspected AC-source. When AC

voltage is present, the indicator [16] will light up. The stronger the AC-

source, the more the indicator [16] will be lit.

7. Battery

•

When the low battery indication (

)

appears, replace the internal

batteries. Turn the rotary switch [3] to the OFF-position.

•

Always disconnect test leads when replacing the battery. Do not us the

device without batteries installed.

•

The battery cover is located on the back of the device and is closed with two

screws. Remove the screws and open the battery compartment.

•

Remove the battery and insert a new 9V (E-block) battery following the

polarity as indicated in the battery holder.

•

Reinstall the cover and secure it with the two screws.

•

Remove the battery when the device is not in use for a longer period of time to

avoid leakage.

•

Do not attempt to recharge non-rechargeable batteries, do not puncture and do

not throw in fire as they may explode.

WARNING: handle batteries with care, observe warnings on

battery casing. Dispose of batteries in accordance with local

regulations.

Keep batteries away from children.

8. Fuses

•

To replace an internal fuse, turn the rotary switch [3] to the OFF-position.

•

Always disconnect test leads when replacing fuses. Do not use the device

without the proper fuses installed.

•

The access the fuses, the whole back cover must be removed. Loosen the 6

outer screws (2 are located under the foldable stand). Do not remove the

battery cover screws (see §7). Lift the whole back cover. Do not touch the

electronic circuitry inside.

•

Replace fuses only by new ones with following specifications:

•

Fuse 1 (large): F 10A/600V fast blow

•

Fuse 2 (small): F 500mA/660V fast blow

Using wrong fuses or short-circuiting fuse holders can lead to

potentially life-threatening situations.

•

Close the back cover and secure the 6 screws.