Appendix c – technical specifications – Velleman LCR40 User Manual

Page 17

Advertising
background image

Atlas LCR User Guide

January 2009 – Rev 4

Page 17

Appendix C – Technical Specifications

Parameter

Min

Typ

Max

Note

range

1Ω

2MΩ

resolution

0.3 Ω

0.6Ω

Resistance

accuracy

Typically ±1.0% ±1.2Ω

1,2,6

range

0.5pF

10,000µF

resolution

0.2pF

0.5pF

Capacitance

accuracy

Typically ±1.5% ±1.0pF

1,2,5

range

1µH

10H

resolution

0.4µH

0.8µH

Inductance

accuracy

Typically ±1.5% ±1.6µH

1,2,4

Peak test voltage (across O/C)

-1.05V

+1.05V

Peak test current (thru S/C)

-3.25mA

+3.25mA

1kHz

-1.5%

±1%

+1.5%

14.925kHz

-1.5%

±1%

+1.5%

Test frequency

accuracy

200kHz

-1.5%

±1%

+1.5%

Sine purity

Typically -60dB 3

rd

harmonic

Operating temperature range

10°C

40°C

3

Battery operating voltage

8.5V

13V


Notes:

1.

Within 12 months of factory calibration. Please contact us if you
require a full re-calibration and/or certification of traceable
calibration.

2.

Specified at temperatures between 15°C and 30°C.

3.

Subject to acceptable LCD visibility.

4.

For inductances between 100µH and 100mH.

5.

For capacitances between 200pF and 500nF.

6.

For resistances between 10Ω and 1MΩ.