Appendix D – Troubleshooting

Problem

Possible Solution

Capacitance measured when

probes are open circuit is not

close to zero (±1.0pF).

Perform a probe compensation.

Resistance and/or inductance

measured when probes are

short circuit is not close to

zero (±1.2Ω, ±1.6µH).

Perform a probe compensation.

Ensure probes are well connected to the

component under test for the entire duration

of the analysis.

Ensure that nothing else is connected with

the component under test. Make sure that

you are not touching the connections.

The component value may be outside the

supported measurement range.

Measured value doesn’t

appear to be correct.

The component’s design frequency may not

correspond to the test frequencies used by

the Atlas LCR.

Measured values vary

slightly between tests.

The displayed resolution is higher than the

measurement resolution to avoid rounding

errors. Small variations within the quoted

measurement resolutions are normal.

Calibration date is

approaching or has past.

Don’t worry, the Atlas LCR will carry on
working after the “Calibration Due Date”

has past. The date is simply a

recommendation.

Probe has become detached.

The probes simply fit to the test lead using

square-pin connectors, carefully push them

back together.