Testing Inductors


The Atlas LCR is designed to analyse most inductors, coils and
chokes.

Inductor test frequency

: The test frequency that the Atlas LCR uses will be

automatically selected from 1kHz, 15kHz or 200kHz. The following table
shows the test frequencies used for various inductance ranges:

Inductance Range

Test Frequency Used

Between 0µH and 0.3mH

200kHz

Between 0.3mH and 4mH

15kHz

Between 4mH and 10H

1kHz


The inductance range for each test frequency shown in the table above is
approximate

. Effects such as DC resistance, hysteresis and Q factor can

influence the frequency that the Atlas LCR selects for your particular
inductor.

Inductance range

: Values ranging from about 1µH to 10H can be measured,

with a minimum resolution of 0.4µH. The DC resistance of the inductor is
measured from 0.5Ω to 1kΩ with a
minimum resolution of 0.3Ω.

Inductor results:

Following analysis,

the inductance is displayed.

Press

scroll-off

to

display

the

frequency at which the inductance was
measured.

Pressing the scroll-off button again will
display the inductor's DC resistance.

Inductance measured for some components can be dependant on the test

frequency used. The effect of frequency on inductance varies depending on
the type of windings and core utilised. Even air cored inductors can show
significant changes of measured inductance at different frequencies.

Inductance

Inductance

Inductance

Inductance

1.507mH

1.507mH

1.507mH

1.507mH

Test frequency

Test frequency

Test frequency

Test frequency

15kHz

15kHz

15kHz

15kHz

DC Resistance

DC Resistance

DC Resistance

DC Resistance

67.2 ohms

67.2 ohms

67.2 ohms

67.2 ohms