Resetting meter settings
This function resets to the default condition meter settings. The
relevant values are given on the following pages:
The following settings are also reset to the default condition:
Resetting the meter
settings
System settings
Measurement settings
Setting
Default settings
Measured variable
ӕ
mS/cm or µS/cm
Adjusted cell constant
0,650 1/cm
Temperature compensation
nLF
Temperature coefficient of the
linear temperature
compensation
2.000 %/K
TDS factor
1.00
1
Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2
During the display test, press <MODE> to open the menu for
the reset of the meter settings.
Init is displayed.
3
Press <
▲><▼> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.