What to do if...

55

ba75558e03

11/2009

6

What to do if...

Error message

E1

Error message,

E3

Display of Cln

Display,

LoBat

Display

to

Instrument does not

react to keystroke

Cause

Remedy

– Measured value outside the

measuring range

– Use suitable measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Unsuitable calibration

solution

– Check calibration solutions

Cause

Remedy

– Cleaning interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Batteries almost empty

– Replace the batteries (see

page 53)

Cause

Remedy

– Time-out of the interface

– Check the instrument that is

connected

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press and hold the <SC>
key and switch the meter on