Xylem CG 853 User Manual
Page 28

Operation
102
4.2.4 Transmitting measured values
You can transmit measured values (data records) in 3 ways:
l
Switch on the data transmission (Int 2) (see page 125)
– After expiry of the selected interval, the current data
record is sent to the interface.
l
Switch on AutoStore (Int 1) (see page 116)
– After expiry of the selected interval, the current data
record is sent to the interface and in addition is stored
in the data store of the instrument
– AutoStore (Int 1) covers the
data transmission
interval
(Int 2).
l
Press the <run/enter> key
This manually transmits the current measured values at
any time - independently of the selected intervals.
Note
If you connect a recorder (analog output), the output to the
digital output is switched off.
7 Perform an AutoRead measurement according to
section 4.2.