Operation

HandyLab 680

50

ba77041e01

03/2014

Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.

6

Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.

7

Wait for the end of the measurement with stability control or
accept the calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).

8

If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.