Xylem HandyLab 680 User Manual
Page 55

HandyLab 680
Operation
55
ba77041e01
03/2014
Continuing with two-
point calibration
11
Thoroughly rinse the IDS sensor with deionized water.
12
Immerse the IDS pH sensor in buffer solution 2.
13
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <
><>.
14
Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
15
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <MENU/ENTER>.
The pH value of the buffer solution is displayed.
16
Set the nominal buffer value for the measured temperature with
<
><>.
17
Accept the calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
18
If necessary, finish the calibration procedure as a two-point cal-
ibration with <M>.
The calibration record is displayed.
15.03.2014 08:00
VariCal