Xylem HandyLab 680 User Manual
Page 54

Operation
HandyLab 680
54
ba77041e01
03/2014
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
7
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <MENU/ENTER>.
The pH value of the buffer solution is displayed.
8
Set the nominal buffer value for the measured temperature with
<
><>.
9
Accept the calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).
10
If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.