Lc l e – Xylem handylab LF 12 User Manual
Page 23

handylab LF 12
Operation
23
ba75367e04
01/2011
4.3.1
Determining the cell constant
(calibrating in the control standard)
Note
This method of automatically determining the cell constant by calibration with
the 0.01 mol/l KCL standard solution can only be used for measuring cells
with cell constants in the range 0.450 ... 0.500 cm
-1
or 0.800 ... 1.200 cm
-1
.
This is how you can determine the cell constant:
Display
Cell constant [cm
-1
]
0.450 ... 0.500 cm
-1
0.800 ... 1.200 cm
-1
E3
Eliminate the error according to
chapter 6 W
outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 1.200 cm
-1
1
Press the <CAL> key until CELL appears on the display.
2
Press the <RUN/ENTER> key.
3
Press the <CAL> key repeatedly, until the calibrated cell con-
stant appears on the display: CAL appears on the display.