7 reset, Reset – Xylem handylab LF 12 User Manual

Page 48

Advertising
background image

Operation

handylab LF 12

48

ba75367e04

01/2011

4.7

Reset

You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.

Measurement

parameters

The following measured parameters (

ϰ

InI) are reset to the default condition:

Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.

Configuration

parameters

The following configuration parameters (InI) are reset to the delivery status:

Measuring mode

ϰ

Cell constant

0.475 cm

-1

(calibrated)

0.475 cm

-1

(set up)

Temperature compensation

nLF

Reference temperature

Temperature coefficient of the lin-
ear temperature compensation

2.000 %/K

TDS factor

1.00

Baud rate

4800

Interval 1
(automatic save)

OFF

Interval 2
(for data transmission)

OFF