Principles of operation, Calibration and effect of temperature, 1 ph – YSI 60 User Manual

Page 23

Advertising
background image

19

8. Principles of Operation

8.1 pH

The YSI Model 60

employs a field replaceable pH sensor for the determination of hydrogen ion

concentration. The sensor is a combination electrode consisting of a proton selective glass
reservoir filled with buffer at approximately pH 7 and a Ag/AgCl reference electrode which
utilizes gelled electrolyte. A silver wire coated with AgCl is immersed in the buffer reservoir.
Protons (H+ ions) on both sides of the glass (media and buffer reservoir) selectively interact with
the glass, setting up a potential gradient across the glass membrane. Since the hydrogen ion
concentration in the internal buffer solution is invariant, this potential difference, determined
relative to the Ag/AgCl reference electrode, is proportional to the pH of the media.

Our testing of the Model 60 pH sensor indicates that it should provide long life, good response
time and accurate readings in most environmental waters, including fresh water of low ionic
strength. No special sensor is required (nor offered) for water of low conductivity.

Calibration And Effect Of Temperature

The software of the YSI Model 60

calculates pH from the established linear relationship between

pH and the millivolt output as defined by a variation of the Nernst equation:

E = E

o

+ 2.3RT * pH

where E = millivolts output

nF

E

o

= a constant associated with the reference electrode

T = temperature of measurement in degrees Kelvin

R, n, and F are invariant constants

Thus, in simplified y = mx + b form, it is (mv output) = (slope)x(pH) + (intercept). In order to
quantify this simple relationship, the instrument must be calibrated properly using buffers of
known pH values. In this procedure, the millivolt values for two standard buffer solutions are
experimentally established and used by the YSI Model 60

software to calculate the slope and

intercept of the plot of millivolts vs. pH. Once this calibration procedure has been carried out,
the millivolt output of the probe in any media can readily be converted by the YSI Model 60
software into a pH value, as long as the calibration and the reading are carried out at the same
temperature.
This last qualifier is almost never met in actual environmental measurements, thus,
a mechanism must be in place to compensate for temperature or, in other words, to accurately
convert the slope and intercept of the plot of pH vs. millivolts established at T

c

(temperature of

calibration) into a slope and intercept at T

m

(temperature of measurement). Fortunately, the

Nernst equation provides a basis for this conversion.

Advertising