Symptom

Possible

Cause

Action

11. Main display reads “PErr”

•

Incorrect sequence of keystrokes.

•

Refer to manual section for step by
step instruction for the function you
are attempting.

12. Main display reads “Err”

(Secondary display reads “ra”)

•

System has failed its RAM test
check procedure.

•

Turn instrument OFF and back ON
again.

•

Return the system for service (pg 21)

13. Main display reads “Err”

(Secondary display reads “ro”)

•

System has failed its ROM test
check procedure.

•

Turn instrument OFF and back ON
again.

•

Return the system for service (pg 21)

16. Main display reads “FAIL”

(Secondary display reads “eep”)

•

EEPROM has failed to respond in
time.

•

Return the system for service (pg 21)

16. Readings on main display don’t

change

•

Meter is in recall mode.

•

Press MODE key to return to Normal
Operation (pg 5, 11)

Advertising