Symptom

Possible

Cause

Action

11. Main display reads “PErr”

Incorrect sequence of keystrokes.

Refer to manual section for step by
step instruction for the function you
are attempting.

12. Main display reads “Err”

(Secondary display reads “ra”)

System has failed its RAM test
check procedure.

Turn instrument OFF and back ON
again.

Return the system for service (pg 21)

13. Main display reads “Err”

(Secondary display reads “ro”)

System has failed its ROM test
check procedure.

Turn instrument OFF and back ON
again.

Return the system for service (pg 21)

16. Main display reads “FAIL”

(Secondary display reads “eep”)

EEPROM has failed to respond in
time.

Return the system for service (pg 21)

16. Readings on main display don’t

change

Meter is in recall mode.

Press MODE key to return to Normal
Operation (pg 5, 11)