For best results:

Calibrate as close as possible to the sample temperature.

After storage in pH 4 buffer/KCl solution, place the pH sensor in pH 7 (6.86) buffer and
allow to acclimate before calibrating (5 to 10 minutes).

Always give the pH and temperature sensors enough time to equilibrate with the
temperature of the buffer.

Temperature sensor

pH sensor (covered by sensor guard)

Figure 5

4. To enter the calibration menu, use two fingers to press and release both the UP ARROW

and DOWN ARROW keys at the same time. The Model 60 display will show CAL at the
bottom, STAND will be flashing and the main display will show 7.00 (the buffer to be used
to adjust the offset).

7.00

pH

STAND CAL

24.8

°C

First buffer value.

Used to adjust the offset.

Flashing

NOTE: If you will be calibrating with pH buffers of 6.86, 4.01 and 9.18 (instead of 7, 4 and
10), press both the UP ARROW and DOWN ARROW keys again. The display will change
to 6.86.

NOTE: The Model 60 automatically accounts for the fact that the true pH of the buffers
changes with temperature, therefore, the pH values displayed during calibration will vary
with temperature. For example, pH 7 buffer at 20°C (rather than 25°C) has an actual pH of
7.02 and this number (rather than 7.00) will appear on the display when the probe is placed
in the solution. See Appendix C - pH Buffer Values.