Troubleshooting, Error messages, Troubleshooting for a list of error messages – YSI 60 User Manual

Page 21

Advertising
background image

17

7. Troubleshooting

Error Messages

The instrument performs a Power On Self Test each time it is turned on. The following error
messages are provided to facilitate troubleshooting. They appear on the LCD when an error is
detected.

Symptom

Possible

Cause

Action

1.

Instrument will not turn on

•

Low battery voltage

•

Batteries installed wrong

•

Meter requires service

•

Replace batteries (pg 2)

•

Check battery polarity (pg 2)

•

Return system for service (pg 21)

16. Instrument "locks up"

•

Instrument has received a shock

•

Batteries are low or damaged

•

System requires service

•

Remove battery lid, wait 15 seconds
for reset, replace lid. (pg 2)

•

Replace batteries (pg 2)

•

Return system for service (pg 21)

16. Instrument will not calibrate due to

unstable readings (decimal point
keeps flashing)

•

pH sensor is fouled

•

pH sensor is bad

•

System requires service

•

Clean pH sensor (pg 14)

•

Replace pH sensor (pg 4, 25)

•

Return system for service (pg 21)

16. pH readings are inaccurate

•

pH buffers out of spec

•

Cal procedure not correct

•

Sample temperature is over 20°C
from cal temperature

•

pH sensor is fouled or damaged

•

pH Sensor is bad

•

System requires service

•

Recalibrate with known good
standards (pg 6)

•

Calibrate within

±20°C of sample

temp (

±10°C for best results)

•

Clean pH sensor (pg 14)

•

Replace pH sensor (pg 4, 25)

•

Return system for service (pg 21)

16. LCD displays "LO BAT"

•

Batteries are low or damaged

•

Replace batteries (pg 2)

16. Main Display reads “nOnE”

•

During recall, no data is currently
stored in memory.

•

Store data before attempting to recall
(pg 11)

16. Main Display reads “OVEr”

•

When calibrating, pH level is over
range allowed for the buffer value
selected.

•

When measuring, pH level is > 14

•

Recalibrate with known good
standards (pg 6)

•

Clean pH sensor (pg 14)

•

Replace pH sensor (pg 4, 25)

•

Return system for service (pg 21)

16. Main Display reads “undr”

•

When calibrating, pH level is under
range allowed for the buffer value
selected.

•

When measuring, pH level is < 0

•

Recalibrate with known good
standards (pg 6)

•

Clean pH sensor (pg 14)

•

Replace pH sensor (pg 4, 25)

•

Return system for service (pg 21)

16. Main Display reads “OVEr”

(Secondary display reads “ovr”)

•

Temperature reading is >75°C

•

Measure samples at a temperature
within the range of the system.

16. Main Display reads “undr”

(Secondary display reads “udr”)

•

Temperature reading is <-5°C

•

Measure samples at a temperature
within the range of the system.