Troubleshooting, Error messages, Troubleshooting for a list of error messages – YSI 60 User Manual

Page 21

Advertising
background image

17

7. Troubleshooting

Error Messages

The instrument performs a Power On Self Test each time it is turned on. The following error
messages are provided to facilitate troubleshooting. They appear on the LCD when an error is
detected.

Symptom

Possible

Cause

Action

1.

Instrument will not turn on

Low battery voltage

Batteries installed wrong

Meter requires service

Replace batteries (pg 2)

Check battery polarity (pg 2)

Return system for service (pg 21)

16. Instrument "locks up"

Instrument has received a shock

Batteries are low or damaged

System requires service

Remove battery lid, wait 15 seconds
for reset, replace lid. (pg 2)

Replace batteries (pg 2)

Return system for service (pg 21)

16. Instrument will not calibrate due to

unstable readings (decimal point
keeps flashing)

pH sensor is fouled

pH sensor is bad

System requires service

Clean pH sensor (pg 14)

Replace pH sensor (pg 4, 25)

Return system for service (pg 21)

16. pH readings are inaccurate

pH buffers out of spec

Cal procedure not correct

Sample temperature is over 20°C
from cal temperature

pH sensor is fouled or damaged

pH Sensor is bad

System requires service

Recalibrate with known good
standards (pg 6)

Calibrate within

±20°C of sample

temp (

±10°C for best results)

Clean pH sensor (pg 14)

Replace pH sensor (pg 4, 25)

Return system for service (pg 21)

16. LCD displays "LO BAT"

Batteries are low or damaged

Replace batteries (pg 2)

16. Main Display reads “nOnE”

During recall, no data is currently
stored in memory.

Store data before attempting to recall
(pg 11)

16. Main Display reads “OVEr”

When calibrating, pH level is over
range allowed for the buffer value
selected.

When measuring, pH level is > 14

Recalibrate with known good
standards (pg 6)

Clean pH sensor (pg 14)

Replace pH sensor (pg 4, 25)

Return system for service (pg 21)

16. Main Display reads “undr”

When calibrating, pH level is under
range allowed for the buffer value
selected.

When measuring, pH level is < 0

Recalibrate with known good
standards (pg 6)

Clean pH sensor (pg 14)

Replace pH sensor (pg 4, 25)

Return system for service (pg 21)

16. Main Display reads “OVEr”

(Secondary display reads “ovr”)

Temperature reading is >75°C

Measure samples at a temperature
within the range of the system.

16. Main Display reads “undr”

(Secondary display reads “udr”)

Temperature reading is <-5°C

Measure samples at a temperature
within the range of the system.