Methods, Chlorine with powder pack: 0.1 – 8.0 mg/l – YSI 900 Chlorine Colorimeter User Manual

Page 14

Advertising
background image

14

Methods

Chlorine with powder pack: 0.1 – 8.0 mg/l

a) Free Chlorine
Fill a clean vial (10 mm ) with

5 ml of the water sample and perform zero

calibration (see “Operation”).
Add the contents of

two Chlorine Free-DpD/F10 powder pack straight from the

foil into the water sample.
Close the vial tightly with the cap and invert several times to mix the contents (20 sec.).
Place the vial in the sample chamber making sure that the

marks are aligned.

Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.

b) total Chlorine
Fill a clean vial (10 mm ) with

5 ml of the water sample and perform zero

calibration (see “Operation”).
Add the contents of

two Chlorine total-DpD/F10 powder pack straight from the

foil into the water sample.
Close the vial tightly with the cap and invert several times to mix the contents (20 sec.).
Place the vial in the sample chamber making sure that the

marks are aligned.

Wait for a reaction period of 3 – 6 minutes.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.

c) Combined Chlorine
combined Chlorine = total Chlorine – free Chlorine

Tolerances:
2 – 3 mg/l: ± 0.2 mg/l

> 3 – 4 mg/l: ± 0.3 mg/l

> 4 – 8 mg/l: ± 0.4 mg/l