YSI 900 Chlorine Colorimeter User Manual

Page 19

Advertising
background image

19

ENGLISH

FaCtOrY CaliBratiOn reset

cAL

SEL

SEL

CAL

CAL

SEL

Resetting the user calibration to the original factory calibration will reset all meth-

ods and ranges.
A user calibrated method is indicated by a "Cal" symbol while the test result is

displayed.
To reset the calibration press both the [MODE] and [ZERO/TEST] key

and

hold.

Turn the unit on using the [ON/OFF] key.

Release the [MODE] and [ZERO/TEST] keys after approx. 1 second.

The following messages will appear will flash on the display:

The factory setting is active.

(SEL stands for Select)

or:

Calibration has been set by the user.

(If the user calibration is to be retained, turn the unit off using the

[ON/OFF] key).

Calibration is reset to the factory setting by pressing the [MODE] key.
The following messages will appear in turn on the display:

CAL for factory calibration; cAL for user calibration.

Turn the unit off using the [ON/OFF] key.

Mode

On

Off

Mode

On

Off

Zero

Test

RESULT

Cal

Store Date
Print Time