YSI 900 Chlorine Colorimeter User Manual
Page 19
19
ENGLISH
FaCtOrY CaliBratiOn reset
cAL
SEL
SEL
CAL
CAL
SEL
Resetting the user calibration to the original factory calibration will reset all meth-
ods and ranges.
A user calibrated method is indicated by a "Cal" symbol while the test result is
displayed.
To reset the calibration press both the [MODE] and [ZERO/TEST] key
and
hold.
Turn the unit on using the [ON/OFF] key.
Release the [MODE] and [ZERO/TEST] keys after approx. 1 second.
The following messages will appear will flash on the display:
The factory setting is active.
(SEL stands for Select)
or:
Calibration has been set by the user.
(If the user calibration is to be retained, turn the unit off using the
[ON/OFF] key).
Calibration is reset to the factory setting by pressing the [MODE] key.
The following messages will appear in turn on the display:
CAL for factory calibration; cAL for user calibration.
Turn the unit off using the [ON/OFF] key.