Basic principles of operation

CR 3200

18

ba76120e01

08/2012

4.2

Operating modes

The thermoreactor has three operating modes:

 Standby

The display shows the model and version number of the
thermoreactor.
Using the g and e keys simultaneously takes you to the SETUP
menu. There you can:

– Edit the 8 temperature programs and the temperature test

program (setting the temperature and reaction time for the
programs 9 to 16 and T, see section 5.5.1)

– Set the display contrast (CONTRAST:0 to 9, see section 5.5.2)

– Activate a manual confirmation before the timer for the reaction

time is started (START TIMER:MAN. or AUTO, see section
5.5.3)

 Program selection

After switching on with e the display shows the current
temperature value for the thermoblock. The second display line
shows the temperature programs and the temperature test program
to be selected if a reaction time of at least one minute has been set
(see section 5.5.1). The control lamps above he display light up
green.

 Active temperature program

The display shows the current temperature value for the
thermoblock. The control lamps above the display light up or flash
red.