5 settings, 1 editing a temperature program, Settings 5.5.1 – YSI CR 3200 Thermoreactor User Manual

Page 25: Editing a temperature program, See section 5.5.1), Section 5.5.1)

Advertising
background image

CR 3200

Operation

25

ba76120e01

08/2012

5.5

Settings

5.5.1

Editing a temperature program

The temperature programs 1 to 8 are installed permanently and cannot
be changed.

The temperature programs 9 to 16 and the temperature test program
can be edited according to individual requirements.

Note
For the temperature test program, a TFK CR external temperature
probe is required (available as an accessory, see chapter 9
A

CCESSORIES

/O

PTIONS

).

1

Switch to the Standby mode.

2

Keep the g key depressed while you press e to switch to the
SETUP menu.
SETUP and, in the second line, an editable parameter are
displayed.

3

Using 8 2 select a temperature program no. 9 to 16 or the
temperature test program.

4

Using g edit the selected temperature program.
The two parameters, temperature and time, are displayed.
The editing is marked on the display by *.
The selected parameter is marked by an arrow < or >.

5

Using 8 2 change the parameter (e.g. temperature ).

6

Using 4 6 switch to the other parameter.

7

Using 8 2 change the parameter (e.g. reaction time ).

8

Using g confirm the changes.
The marking on the display (*) disappears.