7 what to do if, What to do if, Chapter 7 w – YSI CR 3200 Thermoreactor User Manual

Page 31: 7what to do if

Advertising
background image

CR 3200

What to do if...

31

ba76120e01

08/2012

7

What to do if...

There is nothing on the

display

Bars are displayed

instead of the

temperature (-°C)

Temperature deviation

during the temperature

test program

Cuvette emptied /

thermoblock

contaminated

Cause

Remedy

– The power supply is

interrupted

– Check mains cable and

connections

– Exchange the fuses

– Repair by service department

Cause

Remedy

– With an active

temperature test program:
the signal of the external
temperature probe was
not recognized

– Connect the temperature

probe

– Repair by service department

– In the program selection

mode: internal
temperature probe
defective

– Repair by service department

Cause

Remedy

– Bad thermal contact

between the external
temperature probe and the
thermoblock

– Use original accessories only

– Close the protection cover

during the temperature test

– Contact the service

department

Cause

Remedy

– e. g. leaking cuvette

see section 6.4