Maintenance and electrode exchange

6 - 5

ba76003e01

01/2012

6.3

Exchanging the electrodes

Caution

The sensor can be damaged by dirt and moisture. Each time
before dismantling an electrode carefully clean the area around
the electrodes (section 6.2).
Before mounting an electrode make
sure the area behind the sealing ring of the electrode and the
receptacle are dry and clean. The AmmoLyt

®Plus

700 IQ may only

be submersed when the electrodes or original blind plugs are
mounted.

Use the special socket wrench provided to dismantle an electrode.
Electrodes are installed as described in section 3.4.1 E

QUIPPING

THE

SENSOR

WITH

ELECTRODES

.

Recognizing the

electrode type from

outside

When mounted, the electrodes can be recognized by the following
features:

Note
For the correct storage of the electrodes please follow the instructions
in section 3.3 N

OTES

ON

THE

HANDLING

OF

THE

ELECTRODES

.