Matrix adjustment, check and calibration

AmmoLyt

®Plus

700 IQ

4 - 6

ba76003e01

01/2012

The history of matrix adjustments contains the following information:

Note
If you set the potassium electrode to be displayed as the extra sensor
AmmoLyt+K, the calibration history of the potassium electrode can be
viewed under AmmoLyt+K. The IQ S

ENSOR

N

ET

does not keep an extra

log book for this sensor. Log book messages are displayed with the
main sensor, AmmoLyt+.

If you do not set the potassium electrode to be displayed as an extra
sensor, you can create a log book message with the calibration history
by means of the menu item, Calib. history K in the setting table as
necessary. How to do this is described on page 3-10. The log book
message is given in the log book (message code IC5395).

Date

Date of the matrix adjustment

S

Slope [mV] of the electrode.
The slopes are marked by a star (*) because they were not changed
by the matrix adjustment or reset to the default condition with the
electrode zeroing.

DV

Drift voltage [mV]
0 is displayed after an electrode zeroing.

Ref1

Reference concentration [mg/l]

Ref2

This column is of no importance for the matrix adjustment

K+

Measured or entered potassium concentration [mg/l]

P

Procedure (1 = matrix adjustment)

T

Temperature [°C]

R

Evaluation of the matrix adjustment
+ : Matrix adjustment successful.

The sensor measures with the new adjustment data.

? : Matrix adjustment unsuccessful.

Invalid adjustment data was discarded. Measurement is

continued with the last valid values.