9 technical data, 1 measuring characteristics, Technical data -1 – YSI AmmoLyt Plus 700 IQ User Manual

Page 55: Measuring characteristics -1, Chapter 9 t, Echnical, Data, E (chapter 9 t, 9technical data

Advertising
background image

AmmoLyt

®Plus

700 IQ

Technical data

9 - 1

ba76003e01

01/2012

9

Technical data

9.1

Measuring characteristics

Measuring principle

Potentiometric measurement by means of ion sensitive electrodes.
Modular structure with jointly used reference electrode and ion
sensitive electrodes. Integrated microprocessor electronics, screened
2-wire connection for power and data transmission.

Measured parameters

Measuring ranges and

resolution

Measuring ranges and

resolution,

Potassium

measurement

Selectable procedures

for interfering ions

compensation

Main measured parameter

Ammonium

Secondary measured parameter

Temperature

Compensation parameter

Potassium
(depending on the electrode
equipment)

Measuring
mode

Measuring range

Resolution

NH4-N

0.1 ... 100.0 mg/l
1 ... 1000 mg/l

0.1 mg/l
1 mg/l

NH4

0.1 ... 129.0 mg/l
1 ... 1290 mg/l

0.1 mg/l
1 mg/l

mV

-2000 ... +2000 mV

1 mV

Measuring
mode

Measuring range

Resolution

K

1 ... 1000 mg/l

1 mg/l

mV

-2000 ... +2000 mV

1 mV

Compensation procedures

Description

Automatic

up to 1000 mg/l potassium when
equipped with potassium
electrode

Manual

without potassium electrode by
manual entry of the potassium
concentration

(range 1 ... 1000 mg/l).

Advertising